How to Choose a Microcentrifuge or Small Benchtop Centrifuge for Your Laboratory
27 mins read
How to Choose a Microcentrifuge or Small Benchtop Centrifuge for Your Laboratory

By Donal O’Sullivan, BSc. Reviewed by Michael Anderson, MBA.
Choosing a microcentrifuge or small benchtop centrifuge
should start with the laboratory workflow, not the maximum speed printed on the
specification sheet. The right centrifuge needs to accommodate the tubes and
samples you actually use, provide the required relative centrifugal force
(RCF), offer suitable rotor capacity and configuration, and support any
temperature-control or operational requirements of the method. |
Key Takeaways
- Start
with the sample, workflow and tube format, then determine the
centrifuge specification required.
- Compare
centrifuges using the RCF required by the method, rather than
maximum RPM alone.
- Check
the actual rotor configuration, tube capacity and adapter requirements
rather than assuming that a nominal tube size will fit.
- Consider
throughput as well as capacity: the number of tubes processed per run can
materially affect routine laboratory efficiency.
- Choose
refrigeration when the sample, method or validated procedure requires
temperature control, not simply because a refrigerated model is
available.
- Once the technical requirements are defined, compare controls, footprint, accessories and other practical features before moving to individual models.
Start with the Workflow, Not the Centrifuge
A useful first question is not Which centrifuge should we
buy? but:
What does the laboratory need the centrifuge to do?
That distinction matters because apparently similar benchtop
centrifuges can be designed for quite different workloads. A compact unit
intended primarily for quick spin-downs is not necessarily suitable for a
method requiring higher RCF, controlled run times or larger tube formats.
Conversely, purchasing substantially more centrifuge capability than the
workflow requires may add cost and complexity without improving the result.
Begin by documenting the routine work the instrument will
support:
- What
samples will be centrifuged?
- Which
tube formats are used?
- What
RCF does the method or protocol require?
- How
many tubes normally need to be processed at once?
- Is
temperature control required?
- Does
the sample require a sealed or aerosol-tight rotor, bucket or lid, or any
other defined containment feature?
- Are
timed or repeatable runs important?
- Are
multiple workflows expected to share the instrument?
The answers create a basic user requirement against which
candidate centrifuges can be assessed.
This is particularly useful when a centrifuge will be
shared. The most demanding legitimate workflow may determine part of the
required specification, but that does not mean every maximum specification
should be increased. Requirements should be based on actual laboratory needs.
Match the Centrifuge to the Tube and Sample Format
Tube format is one of the most useful starting points
because it immediately narrows the range of suitable equipment.
A laboratory working predominantly with 0.2mL PCR tubes or
1.5/2.0mL microcentrifuge tubes has different equipment requirements from one
routinely processing 15mL or 50mL conical tubes.
A practical first-pass framework is:
|
Routine format |
Typical equipment consideration |
What to establish before selection |
|
0.2 mL PCR tubes |
Compact centrifuge or suitable microcentrifuge
configuration |
Required RCF, PCR-tube or strip compatibility, rotor
supplied and any adapters required |
|
0.5 mL microtubes |
Microcentrifuge or compatible compact centrifuge |
Required RCF, direct fit or adapter requirement, tubes per
run |
|
1.5/2.0 mL microcentrifuge tubes |
Microcentrifuge |
Required RCF, rotor capacity, throughput and control
requirements |
|
5 mL tubes |
Compatible small centrifuge |
Exact tube dimensions, rotor compatibility, capacity,
adapters and required RCF |
|
15 mL conical tubes |
Small benchtop centrifuge with appropriate rotor |
Required RCF, number of tubes per run, rotor configuration
and tube support |
|
50 mL conical tubes |
Benchtop centrifuge with appropriate capacity |
Required RCF, rotor configuration, number of tubes per run
and available space |
This is a selection framework rather than a statement that
every tube of a given nominal volume is interchangeable. Actual tube
dimensions, construction, rated operating limits and the centrifuge
manufacturer's rotor and adapter specifications still need to be checked.
Where a laboratory has not yet established whether 15 mL or
50 mL tubes best suit its workflow, our guide to choosing between 15mL and 50mL centrifuge tubes addresses that decision
separately.
For molecular biology and PCR applications, tube
requirements may also be application-specific. See Choosing Centrifuge Tubes for Molecular Biology and PCR Workflows for the
consumables side of that decision.
Determine the Required RCF
Once the sample and tube formats are understood, establish
the relative centrifugal force required by the workflow.
This is more useful than selecting a centrifuge primarily by
its maximum revolutions per minute (RPM).
RPM describes rotational speed. RCF expresses the
centrifugal acceleration applied to the sample relative to Earth's
gravitational acceleration and depends on both rotational speed and rotor
radius. Two centrifuges operating at the same RPM can therefore generate
different RCF values.
If a method specifies an RCF, that should normally be the
relevant performance requirement when assessing whether a centrifuge can
perform the method.
The objective is not simply to purchase the instrument with
the largest maximum RCF. Instead ask:
- What
RCF does the intended workflow require?
- Can
the proposed centrifuge and relevant rotor achieve it?
- Can
the intended tubes withstand the required conditions when used in the
specified configuration?
- Does
the instrument provide the level of control required for the workflow?
A centrifuge capable of generating a high RCF does not
automatically make every tube that physically fits the rotor suitable for
operation at that force. Confirm the tube manufacturer's maximum RCF and the
conditions attached to that rating, including the required rotor or carrier
support and any applicable limitations relating to temperature, fill volume,
sample or liquid properties and run duration. Where actual conditions differ
from those used to establish the rating, the published maximum RCF may not
apply.
For a more detailed explanation of these relationships, see Understanding RCF, RPM and Rotor Compatibility for Centrifuge Tubes.
Check Rotor Configuration, Capacity and Adapters
A centrifuge specification cannot be assessed properly
without considering the rotor.
The rotor determines which tubes can be accommodated, how
many can be processed and, together with the instrument, the operating
conditions available to the laboratory.
Rotor geometry may also be method-relevant. Where the
workflow depends on tube orientation, pellet location or preservation of
interfaces during centrifugation, confirm whether a fixed-angle,
swinging-bucket or other specified rotor configuration is required rather than
treating different rotor types as interchangeable solely because they accept
the same tube.
Check the actual configuration being supplied rather than
assuming that every rotor or adapter shown for an instrument is included as
standard.
Questions to ask include:
- Which
rotor is supplied with the centrifuge?
- How
many tubes does it accommodate?
- Which
tube formats and dimensions are supported?
- Are
adapters required?
- If
adapters are needed, are they supplied or purchased separately?
- What
operating limits apply to the relevant rotor and configuration?
- Will
another rotor be needed for a second workflow?
Where samples, spills, cleaning agents or disinfectants may
expose the rotor, adapters, lids or seals to aggressive chemicals, confirm that
the relevant materials are compatible with those substances and with the
laboratory's intended cleaning or decontamination procedure. Chemical exposure
can damage rotor and accessory materials even where the sample tube itself
remains compatible.
This becomes particularly important when one centrifuge is
expected to handle several tube formats.
An instrument may appear flexible because additional rotors
or adapters are available, but the laboratory should distinguish between possible
configurations and the configuration actually being purchased.
Capacity should reflect routine workload
Maximum tube capacity is also not necessarily the same as
appropriate capacity.
Consider a laboratory routinely processing 18 microtubes at
a time. A centrifuge accommodating only a small number per run may technically
perform the separation but require multiple cycles. The operational impact can
become significant when repeated several times each day.
Conversely, a larger-capacity instrument may offer little
benefit if most runs contain only a few samples.
Estimate:
typical tubes per run × expected runs per day
and consider peak workloads where these are genuinely
relevant.
The objective is sufficient practical capacity without
automatically selecting the largest available instrument.
Decide Whether Refrigeration Is Required
Refrigerated centrifuges can be important where temperature
control is necessary for the sample, method or established laboratory
procedure. Refrigeration should therefore be treated as a technical
requirement to establish, not a default upgrade.
Ask:
- Is
the sample temperature-sensitive under the intended centrifugation
conditions?
- Does
the method specify a controlled temperature?
- Is
temperature control required by an established or validated procedure?
- Are
run duration and operating conditions likely to make temperature control
important?
- Is
the laboratory replacing an instrument within an existing controlled
workflow?
Where the answer is no, a non-refrigerated centrifuge may be
entirely appropriate.
Where temperature control is genuinely required, confirm the
relevant instrument's temperature range, temperature-control performance and
applicable operating conditions from authoritative manufacturer documentation
rather than assuming that all refrigerated centrifuges provide equivalent
performance. Do not assume that the displayed or selected chamber temperature
is necessarily the temperature reached by the sample throughout the run; where
sample temperature is method-critical, establish any required rotor pre-cooling
or other manufacturer-specified temperature-control procedure.
This distinction can materially affect both equipment
specification and purchase cost, so it is worth resolving before individual
models are compared.
Consider the Controls Needed for Routine Work
After establishing tube format, RCF, rotor configuration and
temperature requirements, consider how the centrifuge will actually be
operated.
Useful features depend on the workflow rather than on the
length of the specification sheet.
RCF and RPM control or display
Where methods are expressed in RCF, the ability to work
directly with RCF can simplify routine operation. Whatever interface is
provided, users still need to understand the relationship between the method
requirement and the selected rotor configuration.
Timer and continuous operation
A timer is useful where centrifugation duration is
controlled as part of a method. Check that the available time range and
operating modes suit the intended work.
Acceleration and deceleration control
Some methods require controlled acceleration or
deceleration, including reduced braking or no braking, to avoid disturbing
pellets, gradients or separated layers. Where the procedure specifies an
acceleration or braking condition, confirm that the centrifuge provides the
required control rather than assuming that speed and run time alone define the
centrifugation step.
Quick-spin operation
Quick-spin or pulse-type operation can be useful for brief
spin-down workflows, but it should not be confused with the controlled
centrifugation requirements of methods specifying particular force and duration
conditions.
Programmable operation
Stored programmes may be valuable where laboratories
repeatedly use defined settings or where several users share an instrument.
They may offer much less value in a laboratory performing only occasional
straightforward spin-downs.
Safety and imbalance controls
Review the actual safety and monitoring features of
shortlisted instruments and their operating instructions. Do not assume that
similarly positioned centrifuges have identical controls.
The general principle is simple: select controls that
improve safe, repeatable operation of the workflows the laboratory actually
performs.
Consider the Practical Fit in the Laboratory
Technical performance is only part of equipment suitability.
A small benchtop instrument still needs an appropriate
location, electrical supply and sufficient working space. Manufacturer
requirements for installation, ventilation, clearances and operating
environment should be checked before purchase.
Other practical considerations may include:
- available
bench space;
- instrument
dimensions and weight;
- access
around the instrument;
- noise
where the centrifuge will operate close to occupied work areas;
- ease
of rotor or adapter changes where several workflows share the instrument;
- cleaning
requirements;
- availability
of required rotors and adapters;
- manufacturer-specified
inspection, maintenance and replacement requirements for the rotor, seals
and other safety-critical accessories, including any defined service-life
limits;
- documentation
needed by the laboratory;
- warranty
and support arrangements.
For regulated, accredited or validated workflows, also
establish whether the supplier or manufacturer can provide the calibration,
performance verification, qualification services and traceable documentation
required by the laboratory's quality system. These requirements should be
defined before purchase where centrifuge speed, time or temperature performance
forms part of a controlled method.
These factors should generally be assessed after technical
suitability has been established. A smaller footprint or lower purchase price
does not compensate for an instrument that cannot safely perform the required
workflow.
A Practical Centrifuge Selection Framework
Rather than comparing product specifications immediately,
work through the selection in a controlled sequence.
|
Step |
Question |
Selection outcome |
|
1. Workflow |
What samples and processes will the centrifuge support? |
Defines the actual laboratory requirement |
|
2. Tube format |
Which tubes must be accommodated? |
Narrows appropriate centrifuge and rotor configurations |
|
3. Required RCF |
What centrifugal force does the workflow require? |
Establishes the necessary performance range |
|
4. Throughput |
How many tubes need to be processed per run and per day? |
Establishes practical rotor capacity |
|
5. Rotor/configuration |
Which rotor and adapters support those tubes and
conditions? |
Confirms physical and operational suitability |
|
6. Temperature |
Does the workflow require controlled temperature? |
Determines whether refrigeration is required |
|
7. Controls |
Which timing, display, programming or other operational
controls are genuinely useful? |
Defines practical operating requirements |
|
8. Laboratory fit |
Does the instrument fit the available space and
operational environment? |
Confirms practical suitability |
|
9. Commercial comparison |
Which technically suitable options best meet the
laboratory's remaining purchasing requirements? |
Creates a defensible shortlist |
This sequence helps prevent a common purchasing problem:
comparing headline specifications before defining what the laboratory actually
needs.
For laboratories primarily requiring microtube
centrifugation, the next commercial step is to explore laboratory microcentrifuges. Where the required tube capacity or
configuration extends beyond that equipment class, the broader laboratory centrifuges range provides the more appropriate starting point.
Pre-Purchase Centrifuge Specification Check
Before requesting quotations or comparing individual
instruments, record the laboratory's requirements.
|
Requirement |
Laboratory specification |
|
Samples/workflows |
|
|
Required tube formats |
|
|
Required RCF or operating range |
|
|
Typical tubes per run |
|
|
Peak capacity requirement |
|
|
Required rotor configuration |
|
|
Required adapters |
|
|
Refrigeration required? |
|
|
Timing/control requirements |
|
|
Quick-spin requirement |
|
|
Programme requirements |
|
|
Available bench space |
|
|
Required documentation |
|
|
Other laboratory-specific requirements |
This specification does not need to be complicated. Its
purpose is to separate requirements from preferences.
A feature should be treated as essential when the workflow
genuinely depends on it. Other features can then be evaluated as preferences
when technically suitable instruments are compared.
That approach also makes discussions with different
suppliers more consistent because each candidate can be assessed against the
same underlying laboratory requirement.
Choosing the Right Centrifuge: The Main Principle
The most appropriate microcentrifuge or small benchtop
centrifuge is not necessarily the fastest instrument or the model with the
longest feature list.
It is the centrifuge whose tube capacity, rotor
configuration, achievable RCF, temperature capability and operational controls
match the laboratory's actual workflow.
Start by defining those requirements. Then compare
technically suitable instruments against the same specification.
This provides a more reliable basis for purchasing decisions
and reduces the risk of selecting an instrument because of headline
specifications that have little relevance to its intended use.
Explore More
- Laboratory Microcentrifuges — explore equipment for microtube and related
small-volume centrifugation workflows.
- Laboratory Centrifuges — explore the broader centrifuge range where capacity or
workflow requirements extend beyond microcentrifuges.
Frequently Asked Questions
What should I look for when choosing a microcentrifuge?
Start with the tube formats, required RCF and number of
samples you need to process. Then check rotor configuration, capacity, adapter
requirements, temperature requirements and the operational controls needed for
the workflow. Compare individual instruments only after these technical
requirements have been established.
Should I compare microcentrifuges by RPM or RCF?
Use the RCF required by the workflow as the more relevant
performance criterion where the method specifies centrifugal force. RPM alone
is not directly comparable between different rotor configurations because the
RCF generated also depends on rotor radius.
Do I need a refrigerated microcentrifuge?
A refrigerated microcentrifuge is appropriate when the
sample, method or established procedure requires temperature control during
centrifugation. Refrigeration should not automatically be treated as necessary
for every laboratory workflow; establish the actual temperature requirement
before selecting the instrument.
Can the same centrifuge be used with different tube
sizes?
Potentially, but only where the centrifuge, rotor and any
required adapters are specified for those tube formats and operating
conditions. Do not assume that a tube is suitable simply because it can
physically be placed in a rotor or adapter.
How much centrifuge capacity does a laboratory need?
Choose capacity according to realistic routine and peak
workloads rather than maximum capacity alone. Consider the typical number of
tubes per run, expected runs per day and whether multiple workflows will share
the centrifuge. An undersized rotor can create unnecessary repeat runs, while
unused capacity may provide little practical benefit.
Read More
- Understanding RCF, RPM and Rotor Compatibility for Centrifuge Tubes
- 15ml vs 50ml Centrifuge Tubes – How to Choose the Right Volume for Your Workflow
- Choosing Centrifuge Tubes for Molecular Biology and PCR Workflows
Written by: Donal O’Sullivan, BSc, Co-Founder and Sales Director, LabFriend UK. Donal brings deep chemistry-led technical expertise across analytical chemistry, biochemistry, environmental monitoring, laboratory instrumentation, consumables and scientific product selection.
Reviewed by: Michael Anderson, MBA, Founder and Managing Director, LabFriend UK. Michael reviews LabFriend UK content for customer relevance, commercial accuracy, operational practicality and alignment with LabFriend UK’s laboratory supply model.
Frequently asked questions
Delivery times vary by product and supplier. Each product page shows current stock levels and the expected lead time, and we confirm this when you place your order. Most items arrive within 2 to 3 weeks. Specialised items can take longer if they are not in stock. If you need something by a certain date, contact us and we’ll check for you.
Yes. Register for a commercial account and, once it’s approved, you will be able to process orders via PO Once you have a commercial account, you can directly email your purchase orders
Add products to your cart and click Get a Quote. You’ll get a formal quote instantly, ready to send to your email address. For bulk orders or custom requirements, send us your product list and we’ll reply with itemized pricing.
If your item is faulty, not as described or sent in error, let us know within 14 days of delivery and we’ll arrange a return. Change of mind returns are only accepted if the item is unused and in its original packaging. A restocking fee will apply and you pay the return freight. Read our full Returns Policy for details.
Yes, prices are shown including VAT by default. Use the Prices Inc. VAT toggle at the top of the page to view prices without VAT.