Low-retention pipette tips are worth considering when liquid remaining on the inside of a conventional tip could materially affect sample recovery, reagent use or transfer consistency. They can be particularly relevant for low-volume transfers and for liquids that tend to leave a film on the internal tip surface, including some detergent-containing, protein-containing or viscous solutions. They are not automatically the best choice for every pipetting task. For routine aqueous solutions where residual liquid is not significant to the workflow, a suitable conventional pipette tip may be entirely adequate. The practical question is therefore not whether low-retention tips are universally “better”, but whether reduced liquid retention provides a meaningful benefit for the particular liquid, volume and method. |
During aspiration and dispensing, liquid interacts with the
internal surface of the pipette tip. Depending on the properties of the liquid
and the tip surface, some liquid may remain behind after dispensing rather than
being transferred to the receiving vessel.
Low-retention pipette tips are designed to reduce this
residual liquid.
Manufacturers achieve low-retention behaviour using
different approaches to the internal tip surface. Depending on the product,
these may include modification of the polypropylene surface, plasma treatment
or other manufacturing technologies intended to reduce wetting and liquid
retention. Some low-retention surfaces are more hydrophobic than conventional
polypropylene, but the mechanism and resulting surface properties are
product-specific and should be confirmed from the manufacturer’s technical
documentation.
Importantly, “low retention” is not a single standardised
performance specification. Technologies and measured performance can differ
between manufacturers and tip ranges. Laboratories should therefore avoid
assuming that all products described as low-retention will behave identically.
For a broader explanation of pipette tip formats and their
role within liquid handling, see LabFriend UK’s Complete
Guide to Pipette Tips, Pipetting Accuracy & Laboratory Liquid Handling.
A small amount of residual liquid is not necessarily
important in every transfer.
The significance depends on what is being transferred, how
much is being transferred and what happens downstream. The same amount of
residual liquid could be inconsequential during one routine transfer but more
relevant when the working volume is small or the reagent is difficult or
expensive to replace.
Low-retention tips therefore become most useful when the
consequences of incomplete delivery matter to the method.
As transfer volumes decrease, liquid remaining on the tip
surface can become more important relative to the intended transfer.
This does not mean that a low-retention tip automatically
solves low-volume pipetting problems. Pipette operating range, the performance
of the pipette with the selected tip, tip fit and seal, operator technique,
evaporation and the physical properties of the liquid can all influence
low-volume performance. A pipette that has been calibrated or verified with one
tip type should not automatically be assumed to provide identical delivery
performance with a different tip.
However, where visible or otherwise significant liquid
retention is occurring, the tip surface is one variable worth investigating.
Some liquids wet polypropylene surfaces more readily than
water and can leave a film on the internal wall of a conventional tip after
dispensing.
Detergent-containing solutions are a common laboratory
example. Manufacturer studies using solutions containing detergents have
demonstrated that residual liquid can differ between conventional and
low-retention tip surfaces.
The extent of the benefit is dependent on the liquid and the
particular tip technology. It should not be converted into a universal
percentage improvement applicable to every laboratory workflow.
Where reagents are expensive, difficult to prepare or
available only in small quantities, unnecessary residual liquid may have a
practical cost.
A low-retention tip may therefore be worth evaluating where
the laboratory repeatedly observes material remaining in the tip or where
recovery is sufficiently important to justify tighter control of this part of
the transfer.
The objective is not simply to buy a more specialised
consumable. It is to determine whether reducing retention produces a useful
improvement in the actual workflow.
Proteins, enzymes and other biological samples can present
recovery challenges, but their behaviour depends on the particular biomolecule,
formulation, concentration and liquid-handling conditions.
Low-retention tips may therefore be useful in some protein
and enzyme workflows, particularly where sample loss or reagent recovery is a
concern.
Low retention and low protein binding are not
interchangeable specifications. A low-retention tip is designed primarily to
reduce liquid remaining on the tip surface after dispensing; this does not by
itself demonstrate that adsorption of a particular protein, peptide or enzyme
to the tip material is reduced. Where biomolecule adsorption is a material
concern, laboratories should review product-specific protein-binding data or
evaluate recovery using the actual biomolecule and formulation rather than relying
on a low-retention claim alone.
High viscosity can slow liquid movement during aspiration
and dispensing and can increase the time required for complete delivery.
Residual liquid from a viscous sample may therefore reflect both its flow
behaviour and its interaction with the tip surface; a low-retention surface
addresses the latter but does not remove the effects of viscosity itself.
Low-retention tips may help reduce surface-related retention
for some viscous liquids, but they do not correct aspiration or dispensing
errors caused by viscosity. Depending on the liquid and method, slower
aspiration and dispensing, an appropriate dwell time, reverse pipetting, or use
of a positive-displacement pipette may provide better control. The pipette
manufacturer’s instructions should be followed when selecting a technique for
non-aqueous or highly viscous liquids.
A low-retention surface should therefore not be treated as a
substitute for selecting an appropriate technique for a viscous sample.
The choice is best made according to the workflow rather
than by assuming that one tip type should replace the other throughout the
laboratory.
|
Selection consideration |
Conventional pipette tips |
Low-retention pipette tips |
|
Routine aqueous transfers |
Often suitable where residual liquid is not significant |
May offer little practical advantage if retention is
already negligible |
|
Low-volume work |
May be suitable where transfer performance is already
acceptable |
Worth evaluating where residual liquid contributes
materially to the transfer |
|
Detergent-containing or low-surface-tension liquids |
Some liquids may leave a film on the internal surface |
Designed to reduce internal liquid retention |
|
Valuable or limited samples |
Suitable where recovery is satisfactory |
May be useful where reducing residual liquid has practical
value |
|
Protein or enzyme workflows |
Suitability depends on the specific sample and method |
May be worth evaluating where surface-related sample loss
is a concern |
|
Viscous liquids |
Liquid properties may make complete dispensing more
difficult |
May reduce retention, but does not replace appropriate
pipetting technique |
|
Routine consumable cost |
Usually appropriate where specialist performance is
unnecessary |
Additional cost should be justified by a meaningful
workflow benefit |
The important distinction is between possible technical
benefit and necessary specification.
If a conventional tip already provides satisfactory recovery
and transfer performance for a validated routine workflow, changing to a
low-retention tip simply because the technology exists may add cost without
solving a real problem.
Low-retention performance describes one characteristic of a
pipette tip. It should not be used as shorthand for overall tip quality or
suitability.
Low retention should also not be treated as synonymous with
low binding. Low-retention performance concerns liquid remaining in the tip
after dispensing, whereas low-binding performance concerns adsorption of sample
components such as proteins or other biomolecules to the material surface. If
adsorption of a specific analyte is a concern, require evidence relevant to
that analyte rather than assuming that a low-retention claim demonstrates low
binding.
A low-retention tip does not by itself establish:
These requirements need to be assessed independently.
This is particularly important when laboratories are
comparing tip suppliers. A tip can have useful low-retention characteristics
and still be unsuitable for a particular pipette or workflow. Conversely, a
conventional tip with good functional fit may be entirely appropriate where
liquid retention is not a significant source of loss or variability.
A proportionate assessment starts with the workflow rather
than the product label.
1. Identify whether retention is actually relevant.
Check whether liquid consistently remains in the tip after dispensing, but do
not rely on visual inspection alone. Residual films can be difficult to see,
particularly with clear detergent-containing solutions. Where retention could
affect the method, assess its practical effect using a suitable quantitative
comparison, such as gravimetric delivery, photometric measurement with an
appropriate test solution, or recovery of the relevant analyte.
2. Consider the consequence.
The case for low-retention tips is stronger when incomplete delivery could
affect a small-volume transfer, consume valuable reagent or reduce recovery of
limited sample.
3. Separate retention from other causes.
If transfers are inconsistent, do not assume the tip surface is responsible.
Pipette condition, tip fit, aspiration and dispensing technique, evaporation
and liquid properties may also contribute.
4. Compare realistic alternatives.
Where low-retention tips appear justified, compare products using the actual
pipette, working volume and representative liquid wherever practical. Do not
assume that products carrying the same “low retention” description provide
equivalent performance.
5. Confirm the other required specifications.
Check manufacturer documentation for the tip's volume range, pipette
compatibility and any sterility, purity, filter or other specifications
required by the workflow.
For an established or performance-sensitive method, compare
the current and proposed tips using the same pipette model, operator, target
volume, liquid, dispensing technique and environmental conditions. Define the
method-relevant acceptance criteria before testing, use replicate transfers for
each tip type, and compare both average recovery or delivered volume and
transfer-to-transfer variability. Where the method is validated or otherwise
controlled, document the assessment through the laboratory’s applicable
change-control procedure before routine implementation.
Low-retention tips should not become a default specification
where there is no identified need.
For many routine buffer, reagent-preparation and general
liquid-handling tasks, a correctly fitting conventional polypropylene tip can
provide the required performance. If the liquid dispenses satisfactorily,
recovery is adequate and the method is performing consistently, there may be
little reason to introduce a specialist tip.
This is particularly relevant to laboratories trying to
control recurring consumables expenditure. The aim should be to specify
low-retention performance where it contributes useful value, not to
increase the specification of every pipette tip used across the laboratory.
For an established workflow, the purchasing comparison
should consider more than the unit price of the tip. Include any measurable
reduction in reagent or sample loss together with the costs of qualification,
inventory management and implementation of a new consumable. A higher-priced
low-retention tip is justified only where the resulting technical or
operational benefit is relevant to the workflow.
For growing biotechnology laboratories, that can mean
maintaining conventional tips for routine work while evaluating low-retention
formats for selected workflows where reagent recovery or residual liquid is
genuinely important.
The words “low retention” alone are not enough to make a
purchasing decision.
Before selecting a product, check:
This avoids turning “low retention” into an assumed quality
grade. It is a functional characteristic that should be selected alongside the
other specifications required by the application.
Are low-retention pipette tips more accurate than
standard tips?
Low-retention pipette tips are not inherently more accurate than standard pipette tips. They are designed to reduce liquid remaining on the internal tip surface. Where retention causes incomplete delivery, reducing that retention may reduce delivery bias and, in some workflows, improve repeatability. Delivery performance still depends on the complete pipette–tip system, selected volume, liquid properties, pipette condition and operator technique.
Do I need low-retention tips for every low-volume
transfer?
No. Low volume can make residual liquid proportionally more
important, but it does not mean every low-volume workflow requires
low-retention tips. The decision should depend on the liquid, the method and
whether retention is materially affecting the transfer.
Are low-retention tips useful for viscous liquids?
Low-retention pipette tips can be useful for viscous liquids when surface-related liquid retention contributes to incomplete delivery. However, viscosity also affects aspiration and dispensing behaviour, so low-retention tips do not solve every viscosity-related pipetting problem. Appropriate technique and, where necessary, the choice of pipetting system should also be considered.
Should I use low-retention tips for proteins and enzymes?
Low-retention pipette tips are worth evaluating for protein and enzyme workflows when residual liquid could materially affect reagent or sample recovery. Their suitability depends on the particular biomolecule, formulation and workflow. Low retention should not be assumed to reduce adsorption of a protein, peptide or enzyme; where biomolecule adsorption is important, review product-specific binding data or evaluate recovery under representative conditions.
Are all low-retention pipette tips equivalent?
No. “Low retention” is not a single standardised performance
specification, and manufacturers use different technologies and test methods.
Compare relevant documentation and, where performance matters, assess the
proposed tip within the intended workflow.
Can a low-retention tip replace a filter tip?
No. Low retention and aerosol-barrier filtration perform
different functions. A tip may incorporate both characteristics, but
low-retention behaviour alone does not provide aerosol protection.
Low-retention pipette tips are most useful when reducing
residual liquid has a meaningful consequence for the laboratory.
That may be the case with low-volume transfers,
detergent-containing or low-surface-tension liquids, selected viscous samples,
or workflows involving valuable biological reagents. In routine applications
where conventional tips already deliver satisfactory performance, the
additional specification may provide little practical benefit.
The most reliable approach is therefore selective rather
than universal: identify where retention matters, compare appropriate tip
options under representative conditions and confirm that compatibility and
other required specifications are independently satisfied.
Laboratories considering this specification can browse low-retention pipette tip options within the LabFriend pipette-tip range
and compare products for their pipette, volume range and workflow.
Written by: Donal O’Sullivan, BSc, Co-Founder and Sales Director, LabFriend UK. Donal brings deep chemistry-led technical expertise across analytical chemistry, biochemistry, environmental monitoring, laboratory instrumentation, consumables and scientific product selection.
Reviewed by: Michael Anderson, MBA, Founder and Managing Director, LabFriend UK. Michael reviews LabFriend UK content for customer relevance, commercial accuracy, operational practicality and alignment with LabFriend UK’s laboratory supply model.
TheComplete Guide to Pipette Tips, Pipetting Accuracy & Laboratory LiquidHandling